Motorized stage connection
Celly motorized stage setting up instruction
Last updated
Was this helpful?
Celly motorized stage setting up instruction
Last updated
Was this helpful?
Main steps:
Remove manual stage
Connect the focus module
Connect XY-stage
Connect control unit
Initial state.
Remove the condenser.
Unscrew 3 screws at the base of the manual stage to remove it.
Unscrew 4 screws at the stage-holder to remove plastic handle.
Fix the focus module on the focus knob with clamp tie
Push the motor towards the microscope for coupling with the fine focus handle
Do not press too hard, otherwise the grip will deteriorate
Tighten motor screws and close the focus block with the end cap.
Connect the module to the manual stage place using 3 screws.
Move the stage vertically with focus knob so that the distance from the 100 lens to the glass is 3-5 mm
Attach the green sensor to the microscope so that it is located directly under the black strip of the stage.
Using the fine focus knob, lift the table so that the distance between the bar and the sensor is 2-3 mm.
Connect signal cable to the control unit and to the stage.
Connect the power cable and turn on the control unit.
Done! Let's start .