# Motorized stage connection

Main steps:

1. Remove manual stage
2. Connect the focus module
3. Connect XY-stage
4. Connect control unit

### 1. Remove manual stage

![](https://357738419-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2F-MJNQBgYLwFTXBIbt1TJ%2Fuploads%2FikEeDBslutDDuEgZjwPN%2F1-1%20initial.jpeg?alt=media\&token=31c00d71-a15a-4b76-a598-7dd5c751ddb5)

Initial state.

Remove the condenser.

![](https://357738419-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2F-MJNQBgYLwFTXBIbt1TJ%2Fuploads%2FzKUuDOtfT26dVLB4ihuL%2F2%20screws.jpeg?alt=media\&token=bf97af06-709a-4391-b246-26a13c7637db)

Unscrew 3 screws at the base of the manual stage to remove it.

![](https://357738419-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2F-MJNQBgYLwFTXBIbt1TJ%2Fuploads%2FqIp4BQXe7DMDJ5P327U9%2F1%20%D1%80%D0%B0%D0%B7%D0%BE%D0%B1%D1%80%D0%B0%D0%BD.jpg?alt=media\&token=846cccda-bd40-4213-91ab-c2fde392da10)

Unscrew 4 screws at the stage-holder to remove plastic handle.

### 2. Connect the focus module

Fix the focus module on the focus knob with clamp tie

![](https://357738419-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2F-MJNQBgYLwFTXBIbt1TJ%2Fuploads%2F0Gvll1Er9r7YVsO8XwgY%2Fimage.png?alt=media\&token=93814267-38d6-4418-b4ad-db7a7c81090b)

Push the motor towards the microscope for coupling with the fine focus handle

{% hint style="warning" %}
Do not press too hard, otherwise the grip will deteriorate
{% endhint %}

![](https://357738419-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2F-MJNQBgYLwFTXBIbt1TJ%2Fuploads%2Fw8AF6NyYcTc9ImshnIeJ%2Fimage.png?alt=media\&token=8f21c697-daa9-4f07-a51b-c794a094aa0a)

Tighten motor screws and close the focus block with the end cap.

### 3. Connect XY-stage

Connect the module to the manual stage place using 3 screws.&#x20;

{% hint style="info" %}
On  different microscopes screw position can vary - let us know what version of microscope do you use before the stage purchase
{% endhint %}

Move the stage vertically with focus knob so that the distance from the 100 lens to the glass is 3-5 mm

<figure><img src="https://357738419-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2F-MJNQBgYLwFTXBIbt1TJ%2Fuploads%2FCDPBa54goDtwtyIa9pZd%2Fimage.png?alt=media&#x26;token=dffc0279-997f-47aa-a052-14108abf81b6" alt=""><figcaption></figcaption></figure>

Attach the green sensor to the microscope so that it is located directly under the black strip of the stage.

<figure><img src="https://357738419-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2F-MJNQBgYLwFTXBIbt1TJ%2Fuploads%2FAUL6tsZGJHRKnvlO0mG0%2Fimage.png?alt=media&#x26;token=8cf0b765-5533-4158-b761-10b635a1b0e3" alt=""><figcaption></figcaption></figure>

Using the fine focus knob, lift the table so that the distance between the bar and the sensor is 2-3 mm.

### 4. Connect control unit

Connect signal cable to the control unit and to the stage.

&#x20;![](https://357738419-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2F-MJNQBgYLwFTXBIbt1TJ%2Fuploads%2FmUuaeg23ritHSo6dqUvf%2Fimage.png?alt=media\&token=2f25ec31-681e-4b57-93f3-82d08613f407)

Connect the power cable and turn on the control unit. &#x20;

Done! Let's start [automated CBC test](https://docs.celly.ai/supported-tests/blood-wbc-differential).&#x20;