> For the complete documentation index, see [llms.txt](https://docs.celly.ai/llms.txt). Markdown versions of documentation pages are available by appending `.md` to page URLs; this page is available as [Markdown](https://docs.celly.ai/config/stage-1.md).

# Motorized stage connection

Main steps:

1. Remove manual stage
2. Connect the focus module
3. Connect XY-stage
4. Connect control unit

### 1. Remove manual stage

![](/files/xdCGqoOBJLRd5y9V7gta)

Initial state.

Remove the condenser.

![](/files/iKfYqibEuJhAudYpHkBk)

Unscrew 3 screws at the base of the manual stage to remove it.

![](/files/viClXdHtV50MC5zuEyTC)

Unscrew 4 screws at the stage-holder to remove plastic handle.

### 2. Connect the focus module

Fix the focus module on the focus knob with clamp tie

![](/files/tBIWQ0xOKoqxqKcEunwM)

Push the motor towards the microscope for coupling with the fine focus handle

{% hint style="warning" %}
Do not press too hard, otherwise the grip will deteriorate
{% endhint %}

![](/files/1DqmHo3PRoPcqiTuQ20t)

Tighten motor screws and close the focus block with the end cap.

### 3. Connect XY-stage

Connect the module to the manual stage place using 3 screws.&#x20;

{% hint style="info" %}
On  different microscopes screw position can vary - let us know what version of microscope do you use before the stage purchase
{% endhint %}

Move the stage vertically with focus knob so that the distance from the 100 lens to the glass is 3-5 mm

<figure><img src="/files/zIDWODeO3rR0gBPqfR6k" alt=""><figcaption></figcaption></figure>

Attach the green sensor to the microscope so that it is located directly under the black strip of the stage.

<figure><img src="/files/UhcGr9VZtyaTv2C0zFQK" alt=""><figcaption></figcaption></figure>

Using the fine focus knob, lift the table so that the distance between the bar and the sensor is 2-3 mm.

### 4. Connect control unit

Connect signal cable to the control unit and to the stage.

&#x20;![](/files/Q573Vyd5QYI1yCWSQOgo)

Connect the power cable and turn on the control unit. &#x20;

Done! Let's start [automated CBC test](/supported-tests/blood-wbc-differential.md).&#x20;